Software Update Notification


  • HDR-9000 PC type OP software V1.1.02017-09-19 열기
    -Random Chart
    This function can be used when the random setting [ON].
    For temperal usage press Alt button and set RANDOM ON in the function key button.
    Press Alt + Set button or wheel click to randomize the VA chart.
    Support only HDC-9000PF/N after version 1.2.1.

    - Touch Masking
    Now, masking operations can be done by clicking the area of the current chart characters.
    Support only HDC-9000PF/N after version 1.2.1.

    - Control Contrast of the chart
    Press Alt button then the Function key buttons below the test view
    is changed to CONTRAST+, CONTRAST-
    Support only HDC-9000PF/N after version 1.2.1.

    - Log
    HDR-9000 Operation Panel writes the log files of important processes, communications and button pressed.
    Each log files can be exported to USB or specific folder selected.
    fileHDR-9000_installer_1.1.0.zip Download
    fileHDR Quick guide_HDR-9000_PC SW OP.pdf Download
  • HDR-9000 PC type OP software V1.0.102017-05-17 열기
    -FIX: Error when prism notation changing from BDBU/BIBO -> r/theta

    -FIX: the problem that PD value is not set when reveiving data from RK/LM

    -FIX: VA data display problem when the chart type is set to M-type

    -FIX: Correct the VA masking behavior in Program Edit/Custom Edit
    Correct fog value to be changed correctly

    -FIX: fog_val, fog_rg setting item function modification

    -FIX: modify input fusion, input stereo, input coincidence [YES][NO] item behavior during the setting동작 수정
    fileHDR-9000_installer_1.0.10.exe Download
    fileHDR Quick guide_HDR-9000_PC SW OP.pdf Download
  • HDR-9000 PC type OP software V1.0.92016-11-30 열기
    Modifications

    - ADD: chart device support for chinese
    -> CDC-9000, CCP-9000

    - FIX: the action changed when loading the data from LM and RK
    -> Preview without charging lens of transmitted data value and press OK button or SET button to charge lens
    -> Fixed to rotate selected item when mouse wheel is input while moving data list in Load menu

    - FIX: Receiving Kerato data normally
    -> Corrected to display the Kerato Data value of receive data

    - ADD: able to recieve the RK and LM data via CAN Box working with HDR-7000 junction box
    fileHDR-9000_installer_1.0.9.zip Download
    fileHDR Quick guide_HDR-9000_PC SW OP.pdf Download
  • HDR-9000 PC type OP Software V1.0.82016-06-13 열기
    HDR-9000
    - CHANGE: modification VA's size of one of the CDC-4000 C type Number chart.

    Only PC type OP Software
    - ADD: support various Export function
    -> add HDR-Mate Protocol V1.0, V1.1, V1.2(Engineer Mode setting)
    - FIX: incorrect Import function.
    - ADD: when th PC Software starting up, send reset message to Refractor body.
    - ADD: add item on Engineer Mode setting.
    -> Delete File After Import
    -> Auto Export
    -> Export File Name
    -> User Defined Export File Name
    -> Auto Import Path
    fileHDR-9000_installer_1.0.8.zip Download
    fileHDR Quick guide_HDR-9000_PC SW OP.pdf Download
  • HDR-9000 PC type OP Software V1.0.72016-03-21 열기
    HDR-9000
    - FIX: incorrect display of printer footer.
    - FIX: incorrect display of toggle state of [alt] and [shift] buttons in the toolbar.
    - FIX: enable to change visual acuity chart in the default program.
    - FIX: bug fix at chart settings releated to CDC-4000 chart device.
    - CHANGE: modification at user interface.
    - NEW: enable to print result in 21 test mode.
    - NEW: enable to use chart masking functions at custom edit and program edit.
    fileHDR-9000_installer_1.0.7.zip Download
    fileHDR Quick guide_HDR-9000_PC SW OP.pdf Download